Skip to content

NI FGEN Standard Output Functions

Palash Khare edited this page Jun 8, 2021 · 4 revisions

Standard Output Functions

niFgen_ConfigureStandardWaveform

ViStatus niFgen_ConfigureStandardWaveform (ViSession vi, ViConstString channelName, ViInt32 waveform, ViReal64 amplitude, ViReal64 dcOffset, ViReal64 frequency, ViReal64 startPhase);

Purpose

Configures the following attributes of the signal generator that affect standard waveform generation:

  • NIFGEN_ATTRIBUTE_FUNC_WAVEFORM
  • NIFGEN_ATTRIBUTE_FUNC_AMPLITUDE
  • NIFGEN_ATTRIBUTE_FUNC_DC_OFFSET
  • NIFGEN_ATTRIBUTE_FUNC_FREQUENCY
  • NIFGEN_ATTRIBUTE_FUNC_START_PHASE

Note:  You must call the niFgen_ConfigureOutputMode function with the outputMode parameter set to NIFGEN_VAL_OUTPUT_FUNC before calling this function.

Parameters

Input
Name Type Description
vi ViSession Identifies your instrument session. vi is obtained from theniFgen_init, niFgen_InitWithOptions, or niFgen_InitializeWithChannels functions and identifies a particular instrument session.
channelName ViConstString

Specifies the channel name for which you want to configure a standard waveform.

Default Value: "0"

waveform ViInt32

Specifies the standard waveform that you want the signal generator to produce. NI-FGEN sets the NIFGEN_ATTRIBUTE_FUNC_WAVEFORM attribute to this value.

Default Value: NIFGEN_VAL_WFM_SINE

Defined Values

NIFGEN_VAL_WFM_SINE Specifies that the signal generator produces a sinusoid waveform.
NIFGEN_VAL_WFM_SQUARE Specifies that the signal generator produces a square waveform.
NIFGEN_VAL_WFM_TRIANGLE Specifies that the signal generator produces a triangle waveform.
NIFGEN_VAL_WFM_RAMP_UP Specifies that the signal generator produces a positive ramp waveform.
NIFGEN_VAL_WFM_RAMP_DOWN Specifies that the signal generator produces a negative ramp waveform.
NIFGEN_VAL_WFM_DC Specifies that the signal generator produces a constant voltage.
NIFGEN_VAL_WFM_NOISE Specifies that the signal generator produces white noise.
NIFGEN_VAL_WFM_USER Specifies that the signal generator produces a user-defined waveform as defined with the niFgen_DefineUserStandardWaveform function.
Name Type Description
amplitude ViReal64

Specifies the amplitude of the standard waveform that you want the signal generator to produce. This value is the amplitude at the output terminal. NI-FGEN sets the NIFGEN_ATTRIBUTE_FUNC_AMPLITUDE attribute to this value.

For example, to produce a waveform ranging from –5.00 V to +5.00 V, set the amplitude to 10.00 V.

Units: peak-to-peak voltage

Default Value: None

Note:   This parameter does not affect signal generator behavior when you set the waveform parameter of the niFgen_ConfigureStandardWaveform function to NIFGEN_VAL_WFM_DC.

dcOffset ViReal64

Specifies the DC offset of the standard waveform that you want the signal generator to produce. The value is the offset from ground to the center of the waveform you specify with the waveform parameter, observed at the output terminal. For example, to configure a waveform with an amplitude of 10.00 V to range from 0.00 V to +10.00 V, set the dcOffset to 5.00 V. NI-FGEN sets the NIFGEN_ATTRIBUTE_FUNC_DC_OFFSETattribute to this value.

Units: volts

Default Value: None

frequency ViReal64 Specifies the frequency of the standard waveform that you want the signal generator to produce. NI-FGEN sets the NIFGEN_ATTRIBUTE_FUNC_FREQUENCY attribute to this value.

Units: hertz

Default Value: None

Note:  This parameter does not affect signal generator behavior when you set the waveform parameter of the niFgen_ConfigureStandardWaveform function to NIFGEN_VAL_WFM_DC.

Frequency ranges vary from device to device. For information about the frequencies supported on your device, refer to the Features Supported topic for your device in the NI Signal Generators Help.

startPhase ViReal64 Specifies the horizontal offset of the standard waveform that you want the signal generator to produce. Specify this parameter in degrees of one waveform cycle. NI-FGEN sets the NIFGEN_ATTRIBUTE_FUNC_START_PHASE attribute to this value. A start phase of 180 degrees means output generation begins halfway through the waveform. A start phase of 360 degrees offsets the output by an entire waveform cycle, which is identical to a start phase of 0 degrees.

Units: degrees of one cycle

Default Value: 0.00

Note:   This parameter does not affect signal generator behavior when you set the waveform parameter to NIFGEN_VAL_WFM_DC.

Return Value

Name Type Description
status ViStatus Returns the status code of this operation. The status code either indicates success or describes an error or warning condition. You can examine the status code from each call to an NI-FGEN function to determine if an error occurred. To obtain a text description of the status code, call the niFgen_error_message function. To obtain additional information about the error condition, call the niFgen_GetError function. To clear the error information from NI-FGEN, call the niFgen_ClearError function.

The general meaning of the status code is as follows:
Value Meaning
0 Success
Positive Values Warnings
Negative Values Errors

niFgen_DefineUserStandardWaveform

ViStatus niFgen_DefineUserStandardWaveform (ViSession vi, ViConstString channelName, ViInt32 waveformSize, ViReal64[] waveformDataArray);

Purpose

Defines a user waveform for use in either Standard Function or Frequency List output mode.

To select the waveform, set the waveform parameter to NIFGEN_VAL_WFM_USER with either the niFgen_ConfigureStandardWaveform or the niFgen_CreateFreqList function.

The waveform data must be scaled between –1.0 and 1.0. Use the amplitude parameter in the niFgen_ConfigureStandardWaveform function to generate different output voltages.

Note:  You must call the niFgen_ConfigureOutputMode function to set the outputMode parameter to NIFGEN_VAL_OUTPUT_FUNC or NIFGEN_VAL_OUTPUT_FREQ_LIST before calling this function.

Parameters

Input
Name Type Description
vi ViSession Identifies your instrument session. vi is obtained from theniFgen_init, niFgen_InitWithOptions, or niFgen_InitializeWithChannels functions and identifies a particular instrument session.
channelName ViConstString

Specifies the channel name for which you want to define a user standard waveform.

Default Value: "0"

waveformSize ViInt32 Specifies the size of the waveform in samples.

Default Value: 16384
waveformDataArray ViReal64[]

Specifies the array of data you want to use for the new arbitrary waveform. The array must have at least as many elements as the value that you specify in waveformSize.

You must normalize the data points in the array to be between –1.00 and +1.00.

Default Value: None

Return Value

Name Type Description
status ViStatus Returns the status code of this operation. The status code either indicates success or describes an error or warning condition. You can examine the status code from each call to an NI-FGEN function to determine if an error occurred. To obtain a text description of the status code, call the niFgen_error_message function. To obtain additional information about the error condition, call the niFgen_GetError function. To clear the error information from NI-FGEN, call the niFgen_ClearError function.

The general meaning of the status code is as follows:
Value Meaning
0 Success
Positive Values Warnings
Negative Values Errors

niFgen_ClearUserStandardWaveform

ViStatus niFgen_ClearUserStandardWaveform (ViSession vi, ViConstString channelName);

Purpose

Clears the user-defined waveform created by the niFgen_DefineUserStandardWaveform function.

Parameters

Input
Name Type Description
vi ViSession Identifies your instrument session. vi is obtained from theniFgen_init, niFgen_InitWithOptions, or niFgen_InitializeWithChannels functions and identifies a particular instrument session.
channelName ViConstString

Specifies the channel name from which you want to clear a user standard waveform.

Default Value: "0"

Return Value

Name Type Description
status ViStatus Returns the status code of this operation. The status code either indicates success or describes an error or warning condition. You can examine the status code from each call to an NI-FGEN function to determine if an error occurred. To obtain a text description of the status code, call the niFgen_error_message function. To obtain additional information about the error condition, call the niFgen_GetError function. To clear the error information from NI-FGEN, call the niFgen_ClearError function.

The general meaning of the status code is as follows:
Value Meaning
0 Success
Positive Values Warnings
Negative Values Errors

niFgen_ConfigureFrequency

ViStatus niFgen_ConfigureFrequency (ViSession vi, ViConstString channelName, ViReal64 frequency);

Purpose

Configures the frequency of the standard waveform that you want the signal generator to produce.

Parameters

Input
Name Type Description
vi ViSession Identifies your instrument session. vi is obtained from theniFgen_init, niFgen_InitWithOptions, or niFgen_InitializeWithChannels functions and identifies a particular instrument session.
channelName ViConstString

Specifies the channel name for which you want to configure a standard waveform.

Default Value: "0"

frequency ViReal64 Specifies the frequency of the standard waveform that you want the signal generator to produce. NI-FGEN sets the NIFGEN_ATTRIBUTE_FUNC_FREQUENCY attribute to this value.

Units: hertz

Default Value: None

Note:  This parameter does not affect signal generator behavior when you set the waveform parameter of the niFgen_ConfigureStandardWaveform function to NIFGEN_VAL_WFM_DC.

Frequency ranges vary from device to device. For information about the frequencies supported on your device, refer to the Features Supported topic for your device in the NI Signal Generators Help.

Return Value

Name Type Description
status ViStatus Returns the status code of this operation. The status code either indicates success or describes an error or warning condition. You can examine the status code from each call to an NI-FGEN function to determine if an error occurred. To obtain a text description of the status code, call the niFgen_error_message function. To obtain additional information about the error condition, call the niFgen_GetError function. To clear the error information from NI-FGEN, call the niFgen_ClearError function.

The general meaning of the status code is as follows:
Value Meaning
0 Success
Positive Values Warnings
Negative Values Errors

niFgen_ConfigureAmplitude

ViStatus niFgen_ConfigureAmplitude (ViSession vi, ViConstString channelName, ViReal64 amplitude);

Purpose

Configures the amplitude of the standard waveform that you want the signal generator to produce.

Parameters

Input
Name Type Description
vi ViSession Identifies your instrument session. vi is obtained from theniFgen_init, niFgen_InitWithOptions, or niFgen_InitializeWithChannels functions and identifies a particular instrument session.
channelName ViConstString

Specifies the channel name for which you want to configure a standard waveform.

Default Value: "0"

amplitude ViReal64

Specifies the amplitude of the standard waveform that you want the signal generator to produce. This value is the amplitude at the output terminal. NI-FGEN sets the NIFGEN_ATTRIBUTE_FUNC_AMPLITUDE attribute to this value.

For example, to produce a waveform ranging from –5.00 V to +5.00 V, set the amplitude to 10.00 V.

Units: peak-to-peak voltage

Default Value: None

Note:   This parameter does not affect signal generator behavior when you set the waveform parameter of the niFgen_ConfigureStandardWaveform function to NIFGEN_VAL_WFM_DC.

Return Value

Name Type Description
status ViStatus Returns the status code of this operation. The status code either indicates success or describes an error or warning condition. You can examine the status code from each call to an NI-FGEN function to determine if an error occurred. To obtain a text description of the status code, call the niFgen_error_message function. To obtain additional information about the error condition, call the niFgen_GetError function. To clear the error information from NI-FGEN, call the niFgen_ClearError function.

The general meaning of the status code is as follows:
Value Meaning
0 Success
Positive Values Warnings
Negative Values Errors

Table of Contents

Internal Development

Creating and Setting Up a gRPC Server

Server Security Support

Creating a gRPC Client

gRPC Client Examples

Session Utilities API Reference

Driver Documentation

gRPC API Differences From C API

Sharing Driver Sessions Between Clients

C API Docs
NI-DAQmx
NI-DCPOWER
NI-DIGITAL PATTERN DRIVER
NI-DMM
NI-FGEN
NI-FPGA
NI-RFmx Bluetooth
NI-RFmx NR
NI-RFmx WCDMA
NI-RFmx GSM
NI-RFmx CDMA2k
NI-RFmx Instr
NI-RFmx LTE
NI-RFmx SpecAn
NI-RFmx TD-SCDMA
NI-RFmx WLAN
NI-RFSA
NI-RFSG
NI-SCOPE
NI-SWITCH
NI-TCLK
NI-XNET
Clone this wiki locally