Notifications
You must be signed in to change notification settings - Fork 2
Commit
This commit does not belong to any branch on this repository, and may belong to a fork outside of the repository.
Add Ophyd objects for Wiener Crate, Scaler and SR570
- Loading branch information
FXI Operator
committed
Sep 6, 2024
1 parent
014dfd3
commit e600575
Showing
2 changed files
with
168 additions
and
9 deletions.
There are no files selected for viewing
This file contains bidirectional Unicode text that may be interpreted or compiled differently than what appears below. To review, open the file in an editor that reveals hidden Unicode characters.
Learn more about bidirectional Unicode characters
Original file line number | Diff line number | Diff line change |
---|---|---|
@@ -0,0 +1,156 @@ | ||
from ophyd import Device, Component as Cpt, EpicsScaler, EpicsSignal, EpicsSignalRO | ||
from ophyd.device import DynamicDeviceComponent as DDC | ||
from collections import OrderedDict | ||
|
||
|
||
# The source for the scaler code is github.com/NSLS-II-SRX/profile_collection/startup/30-scaler.py | ||
|
||
|
||
class EpicsSignalROLazyier(EpicsSignalRO): | ||
def get(self, *args, timeout=5, **kwargs): | ||
return super().get(*args, timeout=timeout, **kwargs) | ||
|
||
|
||
def _scaler_fields(attr_base, field_base, range_, **kwargs): | ||
defn = OrderedDict() | ||
for i in range_: | ||
attr = "{attr}{i}".format(attr=attr_base, i=i) | ||
suffix = "{field}{i}".format(field=field_base, i=i) | ||
defn[attr] = (EpicsSignalROLazyier, suffix, kwargs) | ||
|
||
return defn | ||
|
||
|
||
class FXIScaler(EpicsScaler): | ||
acquire_mode = Cpt(EpicsSignal, "AcquireMode") | ||
acquiring = Cpt(EpicsSignal, "Acquiring") | ||
asyn = Cpt(EpicsSignal, "Asyn") | ||
channel1_source = Cpt(EpicsSignal, "Channel1Source") | ||
channel_advance = Cpt(EpicsSignal, "ChannelAdvance", string=True) | ||
channels = DDC(_scaler_fields("chan", ".S", range(1, 33))) | ||
client_wait = Cpt(EpicsSignal, "ClientWait") | ||
count_on_start = Cpt(EpicsSignal, "CountOnStart") | ||
current_channel = Cpt(EpicsSignal, "CurrentChannel") | ||
disable_auto_count = Cpt(EpicsSignal, "DisableAutoCount") | ||
do_read_all = Cpt(EpicsSignal, "DoReadAll") | ||
dwell = Cpt(EpicsSignal, "Dwell") | ||
elapsed_real = Cpt(EpicsSignal, "ElapsedReal") | ||
enable_client_wait = Cpt(EpicsSignal, "EnableClientWait") | ||
erase_all = Cpt(EpicsSignal, "EraseAll") | ||
erase_start = Cpt(EpicsSignal, "EraseStart") | ||
firmware = Cpt(EpicsSignal, "Firmware") | ||
hardware_acquiring = Cpt(EpicsSignal, "HardwareAcquiring") | ||
input_mode = Cpt(EpicsSignal, "InputMode") | ||
max_channels = Cpt(EpicsSignal, "MaxChannels") | ||
model = Cpt(EpicsSignal, "Model") | ||
mux_output = Cpt(EpicsSignal, "MUXOutput") | ||
nuse_all = Cpt(EpicsSignal, "NuseAll") | ||
output_mode = Cpt(EpicsSignal, "OutputMode") | ||
output_polarity = Cpt(EpicsSignal, "OutputPolarity") | ||
prescale = Cpt(EpicsSignal, "Prescale") | ||
preset_real = Cpt(EpicsSignal, "PresetReal") | ||
read_all = Cpt(EpicsSignal, "ReadAll") | ||
read_all_once = Cpt(EpicsSignal, "ReadAllOnce") | ||
set_acquiring = Cpt(EpicsSignal, "SetAcquiring") | ||
set_client_wait = Cpt(EpicsSignal, "SetClientWait") | ||
snl_connected = Cpt(EpicsSignal, "SNL_Connected") | ||
software_channel_advance = Cpt(EpicsSignal, "SoftwareChannelAdvance") | ||
count_mode = Cpt(EpicsSignal, ".CONT") | ||
start_all = Cpt(EpicsSignal, "StartAll") | ||
stop_all = Cpt(EpicsSignal, "StopAll") | ||
user_led = Cpt(EpicsSignal, "UserLED") | ||
wfrm = Cpt(EpicsSignal, "Wfrm") | ||
mca1 = Cpt(EpicsSignalRO, "mca1") | ||
mca2 = Cpt(EpicsSignalRO, "mca2") | ||
mca3 = Cpt(EpicsSignalRO, "mca3") | ||
mca4 = Cpt(EpicsSignalRO, "mca4") | ||
|
||
def __init__(self, prefix, **kwargs): | ||
super().__init__(prefix, **kwargs) | ||
self.stage_sigs[self.count_mode] = "OneShot" | ||
|
||
sclr1 = FXIScaler("XF:18IDB-ES{Sclr:1}", name="sclr1") | ||
|
||
|
||
|
||
class SR570(Device): | ||
# SR570 preamps are controlled via one-way RS232 connection. The IOC can keep track only of the | ||
# settings change via EPICS. It does not know the actual settings if the changes are made | ||
# manually using buttons on the hardware unit. | ||
|
||
init = Cpt(EpicsSignal, "init.PROC") | ||
reset = Cpt(EpicsSignal, "reset.PROC") | ||
|
||
sensitivity_num = Cpt(EpicsSignal, "sens_num", string=True) | ||
sensitivity_unit = Cpt(EpicsSignal, "sens_unit", string=True) | ||
|
||
offset_on = Cpt(EpicsSignal, "offset_on", string=True) | ||
offset_sign = Cpt(EpicsSignal, "offset_sign", string=True) | ||
offset_num = Cpt(EpicsSignal, "offset_num", string=True) | ||
offset_unit = Cpt(EpicsSignal, "offset_unit", string=True) | ||
offset_u_put = Cpt(EpicsSignal, "off_u_put", ) | ||
offset_u_tweak = Cpt(EpicsSignal, "offset_u_tweak") | ||
offset_cal = Cpt(EpicsSignal, "offset_cal", string=True) | ||
|
||
bias_put = Cpt(EpicsSignal, "bias_put") | ||
bias_tweak = Cpt(EpicsSignal, "bias_tweak") | ||
bias_on = Cpt(EpicsSignal, "bias_on", string=True) | ||
|
||
filter_type = Cpt(EpicsSignal, "filter_type", string=True) | ||
filter_reset = Cpt(EpicsSignal, "filter_reset.PROC") | ||
filter_low_freq = Cpt(EpicsSignal, "low_freq", string=True) | ||
filter_high_freq = Cpt(EpicsSignal, "high_freq", string=True) | ||
|
||
gain_mode = Cpt(EpicsSignal, "gain_mode", string=True) | ||
invert_on = Cpt(EpicsSignal, "invert_on", string=True) | ||
blank_on = Cpt(EpicsSignal, "blank_on", string=True) | ||
|
||
|
||
class SR570_PREAMPS(Device): | ||
unit1 = Cpt(SR570, "{SR570:1}") | ||
unit2 = Cpt(SR570, "{SR570:2}") | ||
unit3 = Cpt(SR570, "{SR570:3}") | ||
unit4 = Cpt(SR570, "{SR570:4}") | ||
|
||
|
||
sr570_preamps = SR570_PREAMPS("XF:18IDB-CT", name="sr570_preamps") | ||
|
||
|
||
class WienerHVCrateChannel(Device): | ||
|
||
status_dec = Cpt(EpicsSignalRO, "StatusDec") # Bits 0-7 | ||
|
||
# Meanings of the status bits: | ||
# outputOn (0) output channel is on | ||
# outputInhibit(1) external (hardware-)inhibit of the output channel | ||
# outputFailureMinSenseVoltage (2) Sense voltage is too low | ||
# outputFailureMaxSenseVoltage (3) Sense voltage is too high | ||
# outputFailureMaxTerminalVoltage (4) Terminal voltage is too high | ||
# outputFailureMaxCurrent (5) Current is too high | ||
# outputFailureMaxTemperature (6) Heat sink temperature is too high | ||
# outputFailureMaxPower (7) Output power is too high | ||
|
||
switch_on_off = Cpt(EpicsSignal, "Switch") | ||
V_set = Cpt(EpicsSignal, "V-Set") | ||
V_sense = Cpt(EpicsSignalRO, "V-Sense") | ||
I_set_limit = Cpt(EpicsSignal, "I-SetLimit") | ||
I_sense = Cpt(EpicsSignalRO, "I-Sense") | ||
temperature = Cpt(EpicsSignalRO, "Temperature") | ||
V_fall_rate = Cpt(EpicsSignal, "V-FallRate") | ||
V_rise_rate = Cpt(EpicsSignal, "V-RiseRate") | ||
|
||
|
||
class WienerHVCrate(Device): | ||
|
||
u0 = Cpt(WienerHVCrateChannel, "HV:u0}") | ||
u1 = Cpt(WienerHVCrateChannel, "HV:u1}") | ||
u2 = Cpt(WienerHVCrateChannel, "HV:u2}") | ||
u3 = Cpt(WienerHVCrateChannel, "HV:u3}") | ||
u4 = Cpt(WienerHVCrateChannel, "HV:u4}") | ||
u5 = Cpt(WienerHVCrateChannel, "HV:u5}") | ||
u6 = Cpt(WienerHVCrateChannel, "HV:u6}") | ||
u7 = Cpt(WienerHVCrateChannel, "HV:u7}") | ||
|
||
|
||
hv_crate = WienerHVCrate("XF:18IDB-OP{WPS:01-", name="hv_crate") | ||
|
This file contains bidirectional Unicode text that may be interpreted or compiled differently than what appears below. To review, open the file in an editor that reveals hidden Unicode characters.
Learn more about bidirectional Unicode characters