Numeric values are assigned to observations that meet certain automated or human identified criterion. Positive flags are considered passing. The meaning of these identifiers are as follows:
Flag | Description |
---|---|
↓ | General |
2 | Measurement data filled from backup data recording source |
1 | Data failed automatic QAQC but manually passed |
0 | Data passes all QAQC metrics |
-1 | Data manually removed |
↓ | Calibration |
-2 | System flush |
-3 | Invalid identifier |
-4 | Drift between adjacent reference tank measurements exceeds limit |
-5 | Time elapsed between reference tank measurements exceeds limit |
-6 | Reference tank measurements out of range |
↓ | 2B 205 |
-10 | O3_ppb measurement out of range (0, 250) |
-11 | Flow_CCmin flow rate out of range (1800, 3000) |
-12 | Cavity_T_C cell temperature out of range (0, 50) |
-13 | Time_UTC recent outage > 1 hour -- warmup period of 5 mins |
↓ | GPS |
20 | position at vehicle storage |
-21 | Fix_Quality invalid gga fix |
-22 | N_Sat number of satellites < 4 |
-23 | Status invalid rmc status |
-24 | Time_UTC recent outage > 5 mins -- warmup period of 1 min |
↓ | Licor 6262 IRGA |
-30 | CO2d_ppm measurement out of range (0, 3000) |
-31 | Flow_mLmin flow rate out of range (395, 405) |
-32 | Cavity_T_C_IRGA cavity temperature out of range (0, 50) |
-33 | Cavity_P_kPa_IRGA cavity pressure out of range (50, 115) |
-34 | Time_UTC recent outage > 1 hour -- warmup period of 5 mins |
↓ | Licor 7000 IRGA |
-40 | CO2d_ppm measurement out of range (0, 3000) |
-41 | Flow_mLmin flow rate out of range (395, 405) |
-42 | Cavity_T_C_IRGA cavity temperature out of range (0, 55) |
-43 | Cavity_P_kPa_IRGA cavity pressure out of range (50, 115) |
-44 | Time_UTC recent outage > 1 hour -- warmup period of 2 mins |
↓ | Los Gatos Research NO2 |
-50 | NO2_ppb measurement out of range (0, 1000) |
-51 | Cavity_P_torr cavity pressure out of range (296, 300) |
-52 | Cavity_T_C cavity temperature out of range (0, 50) |
-53 | Time_UTC recent out > 1 hour -- warmup period of 5 mins |
↓ | Los Gatos Research UGGA |
-60 | CH4d_ppm measurement out of range (0, 1000) |
-61 | CO2d_ppm measurement out of range (0, 3000) |
-62 | H2O_ppm measurement out of range (0, 30000) |
-63 | Cavity_P_torr cavity pressure out of range (135, 145) |
-64 | Cavity_T_C cavity temperature out of range (5, 45) |
-65 | Time_UTC recent outage > 1 hour -- warmup period of 5 mins |
↓ | Magee AE33 |
-70 | BC6_ngm3 measurement out of range (0, 100000) |
-71 | Flow_Lmin flow rate out of range (4.9, 5.1) |
-72 | Time_UTC recent outage > 1 hour -- warmup period of 5 mins |
↓ | MetOne ES642 |
-80 | PM2.5_ugm3 measurement out of range (0, 100000) |
-81 | Flow_Lmin flow rate out of range (1.9, 2.1) |
-82 | Cavity_RH_pct cavity humidty out of range (0, 50) |
-83 | Status alarm status > 0 |
-84 | Time_UTC recent outage > 1 hour -- warmup period of 5 mins |
↓ | Teledyne T200 |
-90 | NO_ppb measurement out of range (-3, 20000) |
-91 | NO2_ppb measurement out of range (-3, 20000) |
-92 | Flow_CCmin flow rate out of range (350, 600) |
-93 | O3_Flow_CCmin ozone flow rate out of range (50, 150) |
-94 | RCel_Pres_inHgA reaction cell pressure out of range (3, 6) |
-95 | Moly_T_C molybdenum converter temperature out of range (305, 325) |
-96 | PMT_T_C photomultiplier tube temperature out of range (5, 12) |
-97 | Box_T_C operating temperature out of range (5, 50) |
-98 | Time_UTC recent outage > 2 hours -- warmup period of 1 hour |
↓ | Teledyne T300 |
-100 | CO_ppb measurement out of range (0, 100000) |
-101 | Flow_CCmin flow rate out of range (500, 1000) |
-102 | Samp_Pres_inHgA sample pressure out of range (15, 35) |
-103 | Samp_T_C sample temperature out of range (10, 100) |
-104 | Bench_T_C optical bench temperature out of range (46, 50) |
-105 | Wheel_T_C filter wheel temperature out of range (66, 70) |
-106 | Box_T_C operating temperature out of range (5, 50) |
-107 | Time_UTC recent outage > 2 hours -- warmup period of 1 hour |
↓ | Teledyne T400 |
-110 | O3_ppb measurement out of range (0, 10000) |
-111 | Flow_CCmin flow rate out of range (500, 1000) |
-112 | Samp_Pres_inHgA sample pressure out of range (15, 35) |
-113 | Samp_T_C sample temperature out of range (10, 50) |
-114 | Box_T_C operating temperature out of range (5, 50) |
-115 | Time_UTC recent outage > 2 hours -- warmup period of 1 hour |
↓ | Teledyne T500u |
-120 | NO2_ppb measurement out of range (0, 1000) |
-121 | Samp_Pres_inHgA sample pressure out of range (15, 35) |
-122 | Phase_T_C gas phase chemiluminescence chamber temperature out of range (15, 35) |
-123 | Box_T_C operating temperature out of range (5, 50) |
-124 | ARef_L_mm auto reference loss out of range (400, 1100) |
-125 | Time_UTC recent outage > 2 hours -- warmup period of 1 hour |
↓ | TEOM 1400ab |
-130 | PM2.5_ugm3 measurement out of range (0, 300000) |
-131 | Time_UTC recent outage > 1 hour -- warmup period of 5 mins |