Broken EDNA V2 Electronics Board #1 #17

Open
fieldja opened this issue Nov 17, 2023 · 1 comment
Open

Broken EDNA V2 Electronics Board #1 #17

fieldja opened this issue Nov 17, 2023 · 1 comment

Comments

@fieldja
Copy link

fieldja commented Nov 17, 2023

This board seems to work fine except for an issue with the shift registers. When running a hyperflush sequence, valves 7 and 8 activate at the same time, and then skips to valve 9. Finally, at the end, the pump runs as if it is trying to pump but no valve opens.

To me this indicates that somehow valves 7 and 8 are being triggered as the same valve and that all the following valves have shifted down by one, leaving one phantom valve at the end.

I am not sure how to solve this issue.

@fieldja
Copy link
Author

fieldja commented Nov 17, 2023

Part 2:

This second board has a similar issue, however, it skips valve 0, and then once it gets to the second half of the sampler, seems to do two valves at a time, but does not skip valves? (IE overlapping sets of 2 valves open at the same time). Probably a TPIC chip issue. The board had a piece of tape that said "010" on it previously, this is probably related to the TPIC issues.

Sign up for free to join this conversation on GitHub. Already have an account? Sign in to comment