Sampler S4.2-001 History #11
Comments
April 2023-Filters Dissolving in Capsules May 2023 |
April 2023- Flow meter not working ***Update May 2023- Flow meter fixed. |
May 2023- Sampler clogging observed- Very low flow rates- Isolated to lower hydraulics September 2023- Sampler fixed- Problem Isolated to Ball valve- Ball valve replaced and operating as normal |
Oct. 4 2023: Wired ball valve on sampler and resolved a poor flow rate issue that was caused due to air leaking into an improper connection on the ball valve, related to issue mentioned in last comment. Sampler seems to be working as it should at this time. |
Two current issues:
|
Oct. 4th, 2023: Fixed issues above with a software update of the server software on the feather to whatever was on the masterv4.2 branch on this date. Sampler seems to be working properly again. |
Flow meter has stopped working post oak creek silt tests. According to notes from tests, filters were used, and flow meter stopped working and could not be repaired. We suspect that the flow meter is still working but was reassembled incorrectly after cleaning. Flow meter will be retested and this issue will be updated today. |
Riley disassembled and tested the flow meter and determined it was reassembled incorrectly. The flow meter is now working again. |